Input Characteristics | Rated input voltage | 100~240Vac |
Operating range | 90~264Vac | |
Rated input current | 3.0A Max | |
Input frequency | 50~60Hz +/-3Hz | |
Power dissipation(No-load) | <0.21W | |
Output Characteristics | Output voltage | 12~50Vdc(according to customer demand) |
Output current | 0.01~16.0A(according to customer demand) | |
Line regulation | Rated voltage ±5% | |
Load regulation | Rated voltage ±5% | |
Ripple and noise | 120mVp-p Max | |
Turn on delay time | ≤1S | |
Rise time | 100mS Max | |
Hold up time | 10mS Min | |
Over-shoot | Rated voltage 10% Max | |
Efficiency | Meet: DOE VI COC V5 TIER2 | |
Protection Characteristics | Over voltage protection | The power supply shall be hiccup mode when the output is over voltage, and the power supply shall not be damaged |
Over current protection | The output shall hiccup when the over current applied to the output, and shall be self-recovery when the fault condition is removed | |
Short circuit protection | The input power shall decrease when the output is short to GND, the power supply shall not damage, and shall be self-recovery when the fault condition is removed |
Environmental Characteristics | Operating temperature | 0~+40°C |
Non-operating temperature | -20~+80°C | |
Operating humidity | 10%~90%(non condensing) | |
No-operating humidity | 10%~90%(non condensing) | |
Safety Standards | Dielectric strength(Hi-pot) | 1500Vac 10mA 60S or 2121Vac 5mA 60S |
Insulation resistance | 100MΩ min at 500V DC |